Agilent Technologies 1100 Series User Manual page 126

Diode array and multiple wavelength detector sl
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

8
Maintenance
Orientation of Flow Cell Parts
NOTE
Gaskets # 6 and #7 have
different hole diameters
1 - window screw
2 - spring washers
3 - compression washer
4 - window holder
5 - quartz window
6 - gasket (light in)
7 - gasket (light out)
8 - window screw (contains
items 2, 3, 4 and 5)
9 - flow cell body
10 - inlet capillary
11 - outlet capillary
12 - holder
3
Assemble the washers and the window assembly in
correct order.
126
8
7
12
10
4
Correct orientation of spring washers [2] is required.
9
6
5
4
11
1100 Series DAD and MWD User Manual
2
3
1

Advertisement