Agilent Technologies 1100 Series User Manual page 65

Diode array and multiple wavelength detector sl
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

6 Allow the detector to warm up of at least 60 minutes to provide a stable
.
For reproducible chromatography, the detector and lamp should be on for at least one hour.
N O T E
Otherwise the detector baseline may still drift (depending on the environment). See also
section
Figure 34
Table 13
Time [minutes]
17 - 20
27 - 30
37 - 40
47 - 50
57 - 60
1100 Series DAD and MWD User Manual
baseline (see example in
"Wander / Drift Problems Due to Temperature Changes"
Stabilization of Baseline (both lamps turned on at the same time)
Baseline drift after lamp turn on (example from
Figure 34
and
Table
Drift [mAU/hr]
2.6
0.8
0.4
0.2
< 0.2
Using the Detector
13).
in the Service Manual.
Figure
34)
4
65