8
Maintenance
3
Disconnect the flow cell inlet capillary (top) and the waste
tubing (bottom) from the unions.
5
Remove the flow cell while pressing the flow cell holder.
122
4
Loosen the thumb screw and remove the flow cell outlet
capillary (bottom) with the union.
Note:
The label attached to the flow cell provides
information on part number, path length and
maximum pressure.
If you want to replace flow cell parts, see
the Standard or Semi-Micro Flow Cell"
or
"Repairing the Standard or Semi-Micro Flow
Cell"
on page 124.
1100 Series DAD and MWD User Manual
"Repairing
on page 124