Initialization mode selection

The following initialization (init) modes are selectable:
Table 6
Initialization Mode Switches
ON
2
1
3
4
5
Bootp
When the initialization mode "Bootp" is selected, the detector tries to
download the parameters from a Bootp Server. The parameters obtained
become the active parameters immediately. They are not stored to the
non-volatile memory of the detector. Therefore, the parameters are lost with
the next power cycle of the detector.
Figure 8
Bootp & Store
When "Bootp & Store" is selected, the parameters obtained from a Bootp
Server become the active parameters immediately. In addition, they are stored
to the non-volatile memory of the detector. Thus, after a power cycle they are
still available. This enables a kind of "bootp once" configuration of the
detector.
1100 Series DAD and MWD User Manual
SW 6
OFF
OFF
OFF
6
7
8
OFF
Bootp
Server
Bootp (Principle)
SW 7
SW 8
Init Mode
OFF
OFF
Bootp
OFF
ON
Bootp & Store
ON
OFF
Using Stored
ON
ON
Using Default
Active
Parameter
LAN Configuration
35
3