Maintenance
2
After replacing the outlet capillary, fix it handtight first.
Then do a 1/4 turn with a 4-mm wrench.
Video:
Unscrew the cell body from the heat
exchanger and the heat exchanger
from the clamp unit.
132
3
To replace the inlet capillary, use a 4-mm wrench for the
fitting.
Outlet
4
Unscrew the fixing screw and unwrap the inlet capillary
from the grove in the flow cell body.
Screw
Inlet capillary
with heat
exchanger
1100 Series DAD and MWD User Manual