6
Troubleshooting and Diagnostics

Detector Status Indicator

The detector status indicator indicates one of four possible detector
conditions:
• When the status indicator is OFF (and power switch light is on), the
• A green status indicator, indicates the detector is performing an analysis
• A yellow indicator indicates a not-ready condition. The detector is in a
• An error condition is indicated when the status indicator is red. An error
• A red-blinking indicator indicates that the module is in resident mode (e.g.
110
detector is in a prerun condition, and is ready to begin an analysis.
(run mode).
not-ready state when it is waiting for a specific condition to be reached or
completed (for example, immediately after changing a set point), or while a
self-test procedure is running.
condition indicates the detector has detected an internal problem which
affects correct operation of the detector. Usually, an error condition
requires attention (e.g. leak, defective internal components). An error
condition always interrupts the analysis.
during update of main firmware).