Agilent Technologies 1100 Series User Manual page 103

Diode array and multiple wavelength detector sl
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Figure 59
With a UV-visible detector based on a diode array and the correct choice of a
reference wavelength setting, quantitative detection is possible. To suppress
caffeine, the reference wavelength must be set to 282 nm. At this wavelength,
caffeine shows exactly the same absorbance as at 222 nm. When the
absorbance values are subtracted from each another, any indication of the
presence of caffeine is eliminated. In the same way, hydrochlorothiazide can
be suppressed if caffeine is to be quantified. In this case the wavelength is set
to 204 nm and the reference wavelength to 260 nm.
chromatographic results of the peak suppression technique.
The trade-off for this procedure is a loss in sensitivity. The sample signal
decreases by the absorbance at the reference wavelength relative to the signal
wavelength. Sensitivity may be decreased by as much as 10–30 %.
1100 Series DAD and MWD User Manual
WL1 (204 nm, caffeine)
WL2 (222 nm, hydrochlorothiazide
WL3 (260 nm, reference to suppress
hydrochlorothiazide
WL4 (282 nm, reference to suppress caffeine
WL3
WL1
WL2
Wavelength Selection for Peak Suppression
How to optimize the Detector
WL4
Wavelength (nm)
Figure 60
shows the
5
103

Advertisement

Table of Contents
loading

Table of Contents