Maintenance
9
Insert the fixing screw, so that the capillary cannot leave
the grove.
Screw
Inlet capillary
with heat
11
Fix the inlet capillary to the flow cell body handtight first.
Then do a 1/4 turn with a 4-mm wrench.
IN
OUT
134
10
Fix the flow cell body to the heat exchanger.
IN
OUT
12
Fix the new heat exchanger to the clamp unit.
1100 Series DAD and MWD User Manual