Maintenance
3
When replacing the Vis-lamp, assure that the Vis-lamp is
inserted as shown (flat edge towards the deuterium lamp).
5
Replace the front cover.
120
4
Insert the lamp. Fix the screws and reconnect the lamp to
connector.
Next steps:
6
Reset the lamp counter as described in the user interface
documentation (lamps with I.D. tag cannot be reset).
7
Turn the lamp on and give the lamp 10 minutes to warm up.
8
Perform a
"Wavelength Verification and Recalibration"
described in the Service Manual (after UV-lamp
replacement).
9
Perform an
"Intensity Test"
Manual.
1100 Series DAD and MWD User Manual
as described in the Service