With the instrument accessory kit comes a 4-mm wrench and with the Sealing Kit a special
N O T E
adapter. Both together work as a torque wrench with pre-defined torque (maximum allowed
torque for the cell fittings is 0.7 Nm). It can be used to tight the capillary fittings at the flow
cell body. The wrench has to be plugged into the adapter as shown in
Wrench
max
0.7 Nm
Figure 65
1100 Series DAD and MWD User Manual
Wrench plus Torque Adapter