Use a 4 mm hex key to unscrew the window assembly [1]
and remove the gasket [2] from the cell body.
1
2
2
Use a tooth pick to remove the quartz window from the
window assembly (push through from the screw side).
1100 Series DAD and MWD User Manual
Note:
If you want to replace the gasket only, continue with
step 7.
Do not mix the gasket #6 and # 7. They are different
for standard and semi-micro flow cell.
2
1
Note:
If the washers fall out of the window assembly, they
must be inserted in the correct order with the
Teflon® ring to prevent any leaks from the flow cell
window.
Maintenance
8