Agilent Technologies 1100 Series User Manual page 38

Diode array and multiple wavelength detector sl
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

3
LAN Configuration
The user may open a Telnet session using the default IP address and change
the parameters stored in the non-volatile memory of the detector. He may then
close the session, select the initialization mode "Using Stored", power-on again
and establish the TCP/IP connection using the new parameters.
When the detector is wired to the PC directly (e.g. using a cross-over cable or a
local hub), separated from the local area network, the user may simply keep
the default parameters to establish the TCP/IP connection.
In the "Using Default" mode, the parameters stored in the memory of the detector are not
N O T E
cleared automatically. If not changed by the user, they are still available, when switching
back to the mode "Using Stored".
38
1100 Series DAD and MWD User Manual