Next steps:
13
Reconnect the capillaries, see
14
Perform a leak test.
15
Insert the flow cell.
16
Replace the front cover.
17
Perform a
"Wavelength Verification and Recalibration"
check the correct positioning of the flow cell.
1100 Series DAD and MWD User Manual
"Exchanging a Flow Cell"
or
on page 121.
"Holmium Oxide Test"
as described in the Service Manual to
Maintenance