How to optimize the Detector
Table 17
Optimization Overview, continued
Parameter
5 Setting the slit width
Use 4 nm slit for normal applications.
Use narrow slit (e.g 1 nm) if your analytes have narrow absorbance
bands and for high concentrations.
Use a wide slit (e.g. 16 nm) to detect very low concentrations.
Optimizing spectral acquisition (DAD only)
Select spectra acquisition mode according to your needs (see
"Spectrum Settings"
Set the spectral wavelength range (for colorless samples 190...400 nm
is sufficient).
Set step to 4 nm for normal use; set small step (and slit width) if high
resolution of spectra with fine structure is wanted.
Typical column
Typical peak
length
width
T<= 5 cm
0.025 min
10 cm
0.05 min
20 cm
0.1 min
>= 40 cm
0.2 min
Typical flow
rate
Internal column diameter
Figure 51
90
on page 81).
Recommended flow cell
Nano
flow cell
0.01 ... 0.2
ml/min
0.5 ... 1 mm
Choosing a Flow Cell
Impact
• spectral resolution, sensitivity and
Semi-mirco
flow cell
Standard
flow cell
0.2 ... 0.4
0.4 ... 0.4
ml/min
ml/min
2.1 mm
3.0 mm
1100 Series DAD and MWD User Manual
linearity.
1 ... 2 ml/min
4.6 mm
High pressure
flow cell for
pressures
above 100 bar
0.01 ... 5
ml/min