Agilent Technologies 1100 Series User Manual page 92

Diode array and multiple wavelength detector sl
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
How to optimize the Detector
Figure 52
Traditionally LC analysis with UV detectors is based on comparing
measurements with internal or external standards. To check photometric
accuracy of the Agilent 1100 DAD/MWD it is necessary to have more precise
information on path lengths of the flow cells.
The correct response is:
expected response * correction factor
Please find below the details of the Agilent 1100 DAD/MWD flow cells:
Table 18
Correction factors for Agilent 1100 DAD/MWD flow cells
Flow cell type
Standard flow cell
Semi-micro flow cell
500 Nanoliter flow cell kit
80 Nanoliter flow cell kit
High Pressure flow cell
92
Time (min)
Influence of Cell Path Length on Signal Height
Cell volume
Part number
13 µl
G1315-60022
5 µl
G1315-60011
0.5 µl
G1315-68724
0.08 µl
G1315-68716
1.7 µl
G1315-60015
Analysis of pesticide standard
6-mm optical path length
10-mm optical path length
Path length
Path length
(nominal)
(actual)
10 mm
9.80 ± 0.07 mm
6 mm
5.80 ± 0.07 mm
10 mm
10.00 ± 0.02 mm
6 mm
6.00 ± 0.02 mm
6 mm
5.75 ± 0.07 mm
1100 Series DAD and MWD User Manual
Correction
factor
10/9.8
6/5.8
10/10
6/6
6/5.75

Advertisement

Table of Contents
loading

Table of Contents