Installing the Detector

Installing the Detector
Preparations
Parts required
Before adding a G1315C DAD or G1365C MWD into an existing 1100 system assure that the
N O T E
existing 1100 modules have been updated to firmware revision A.06.xx. Otherwise the
ChemStation B.01.03 or later will not recognize the 1100 modules.
Assure that the CompactFlash Card is installed in the rear of the detector (required for
N O T E
operation).
1 Note the MAC address of the LAN interface (rear of the module, under the
2 Place the detector in the stack or on the bench in a horizontal position.
3 Ensure the line power switch at the front of the detector is OFF.
4 Connect the power cable to the power connector at the rear of the detector.
5 Connect the CAN cable to other Agilent 1100 Series modules.
6 Connect the LAN cable (e.g. from a Agilent ChemStation as controller) to
In multi-detector configurations the LAN of the G1315C DAD or G1365C MWD must be
N O T E
used due to its higher data load.
24
Locate bench space
Provide power connections
Unpack the detector
• Detector
• Power cord
• LAN cable (cross-over or twisted pair network cable)
• for other cables see below and section
Manual.
• ChemStation (B.01.03) and/or Control Module G1323B (B.04.01)
• Other 1100 modules must have firmware A.06.xx
configuration switch, see
Configuration"
on page 31.
the G1315C/G1365C LAN connector.
Figure 3
on page 25). It's required for
1100 Series DAD and MWD User Manual
"Cable Overview" in the Service
"LAN