Slit Width - Agilent Technologies 1100 Series User Manual

Diode array and multiple wavelength detector sl
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Figure 56

Slit Width

The detector has a variable slit at the entrance of the spectrograph. This is an
effective tool to adapt the detector to changing demand of different analytical
problems.
A narrow slit provides spectral resolution for analytes with very fine
structures in the absorbance spectrum. An example of such a spectrum is
benzene. The five main absorbance bands (fingers) are only 2.5 nm wide and
just 6 nm apart from each other.
1100 Series DAD and MWD User Manual
Grad.: 0.02 m KH
PO
/ACN from 12% ACN to 45% ACN in 12 min
2
4
Gradient Analysis of PTH-Amino Acids (1 pmol each), with and without Refer-
ence
Time (min)
How to optimize the Detector
1 pmol each
Wavelength
267 nm
Reference 380 nm
Wavelength
267 nm
No reference
5
97