Figure 10
Using Default
When "Using Default" is selected, the factory default parameters are taken
instead. These parameters enable a TCP/IP connection to the LAN interface
without further configuration, see
Figure 11
Using the default address in your local area network may result in network problems. Take
N O T E
care and change it to a valid address immediately.
Table 7
IP address:
Subnet Mask:
Default Gateway
Since the default IP address is a so-called local address, it will not be routed by
any network device. Thus, the PC and the detector must reside in the same
subnet.
1100 Series DAD and MWD User Manual
Non-Volatile
RAM
Stored
Parameter
Using Stored (Principle)
Default
Parameter
Using Default (Principle)
Using Default Parameters
Active
Parameter
Table
7.
Active
Parameter
192.168.254.11
255.255.255.0
not specified
LAN Configuration