5 Press OK. The parameter are added to the Bootp Manager, see
Figure 15
6 Press Exit Manager and OK to exit the Agilent Bootp Service.
7 Now turn on the module with the detector, wait about 30-60 seconds and
02/03/05 16:33:56 PM
Status: BOOTP Request received at outer most layer
Status: BOOTP Request received from hardware address: 0030D30A0838
Status: found 134.40.27.95 WADI1171:
Status: Host IP Address is: 134.40.29.56
Status: Reply to BOOTP Request has been sent
Status: BOOTP Request finished processing at outer most layer
Figure 16
1100 Series DAD and MWD User Manual
and added to the TabFile, see
Bootp Manager - check your entries
view the LogFile, see
detector with the hardware (MAC) address.
LogFile - the detector has received the parameter
Figure 12
on page 41:.
Figure
16. It should display the request from the
LAN Configuration
Figure 14
43