Installing the New Dryer
1. Insert the replacement dryer into the dryer port on the analyzer. Use the
2. Tighten the VCR fittings until they are finger tight.
3. Install the dryer mounting screws to secure the dryer to the bracket.
4. Tighten both of the VCR fittings so that the nuts are 1/8 of a turn past
5. Place the Dryer Compartment Cover onto the rear of the analyzer and
6. Enter the production code for the new dryer.
7. Power up the analyzer. Wait until the boot-up sequence is complete.
4-16 | 5830 Moisture Analyzer
Minimize the time that the dryer is exposed to the atmosphere after
removing the protective caps. AMETEK recommends that the dryer be
exposed for less than one (1) minute.
new VCR gaskets provided with the dryer in each of the fittings.
finger tight.
secure with the four (4) captured screws.
Press TEST/CONFIG.
•
•
Scroll down to Production Codes and press Enter.
Select Reference Dryer and press Enter. Enter the dryer production
•
code and press Enter.
Press Cancel several times to reverse through the menu and return to
•
the normal operating screen.
PN 583054901, Rev YA