Maintenance of the Electronics Module

The Electronics Module was designed to easily identify and service the boards
and power supplies used in the 5830 Analyzer. These components include:

Removing the Electronics Module From the 5830

Tools Required

Procedure for Removing the Electronics Module

Use Figure 4-9 as a reference for locating the items described in this procedure.
All of the components are now accessible for maintenance or replacement.
PN 583054901, Rev YA
MCU Assembly
Analog Interface Assembly
Power Supply
Fan
Pilot Valve Manifold
Disconnect all power to the analyzer before removing the Electronics
Module.
1/4-inch nut driver
1/16-inch hex wrench
Phillips and slotted screwdrivers
Locate the four (4) Phillips head screws on the Electronics Module.
Using the appropriate tool, remove the mounting hardware.
With all the interconnecting wiring and tubing still connected, remove
the Electronics Module from the 5830 Enclosure by pulling up and slightly
towards the front of the analyzer.
Once you remove the Electronics Module from the 5830 Enclosure, you
can place it on the 5830 Enclosure as a resting fixture. However, take care
not to scratch the 5830 Enclosure with the Electronics Module (a piece
of cardboard can be placed under the Electronics Module to protect the
enclosure).
Maintenance and Troubleshooting | 4-17