Range Key

Use the Range key to define your 4–20 mA output setting.
Use this setting to scale the 4–20 mA output. The 4–20 mA output is propor-
tional to the moisture readings.
Figure 3-5.
Main Menu, Range key.
Maximum Conc.
Minimum Conc.
Example:
To scale the Analog Output to represent a 0–100 PPMV range, enter the follow-
ing settings:
Maximum Conc.:
Minimum Conc.:
PN 583054901, Rev YA
RANGE
Enter the moisture reading equivalent to 20 mA.
Enter the moisture reading equivalent to 4 mA.
100
0
Enter
Maximum Conc
Value
Enter
Minimum Conc
Value