Installing The New Dryer - Ametek 5800 User Manual

Moisture
Table of Contents

Advertisement

Installing the new dryer

Minimize the time that the dryer is exposed to the atmosphere after
!
removing the protective caps. AMETEK recommends that the dryer
be exposed for less than a minute.
CAUTION
1.
Insert the replacement dryer into the dryer port on the analyzer. Use
the new VCR gaskets provided with the dryer in each of the fittings.
2.
Tighten the VCR fittings until they are finger tight.
3
Install the dryer mounting screws to secure the dryer to the bracket.
4.
Tighten both of the VCR fittings so that the nuts are 1/8
finger tight.
5.
Place the dryer compartment cover onto the rear of the analyzer and
secure with the four captured screws.
6.
Enter the production code for the new dryer.
Power up the analyzer. Wait until the boot-up sequence is complete.
-
Press TEST/CONFIG.
Scroll down to "Production Codes" and press Enter.
-
Select "Reference Dryer" and press Enter to enter the production
code. Press Enter.
-
Press Cancel several times to reverse through the menu and re-
turn to the normal operating screen.
of a turn past
th
|
Service & Parts
7-11

Advertisement