Rear Panel Connections and Interface Board
See Figures 2-2 and 2-3 for location of analyzer connections on the Rear Panel
and Interface board. Remove Terminal Cover to access Interface board.
1
2
3
4
5
6
7
8
9
10 RS-485 Terminator
11 RS-485 Serial Port
12 Analog Output
13 Alarm Contacts
2-8 | 5830 Moisture Analyzer
Item
Power Conduit
Signal Conduit
Instrument Air Fitting
RS-232 Serial Port
Exhaust Outlet Fitting
Sample Inlet Fitting
Power connector
Fuse Receptacle
Line Voltage Selector
Function
Cable gland for analyzer power cord.
Cable gland for signal connections.
Input port for Instrument Air used to operate internal
diaphragm valves; 1/4-inch Swagelok fitting.
User serial communications interface. Serial wiring
supplied by user. Observe caution regarding use in
hazardous locations.
Vent port for cell exhaust; 1/4-inch male VCR connec-
tor.
Input port for Sample gas; 1/4-inch male VCR connec-
tor.
Connect hot (1) and neutral (2) to Power Connector
and ground to chassis.
2 A 250 V, time lag, T-type.
Select 120 or 240 Volt line voltage.
RS-485 Terminator switch is on if the analyzer is the
only analyzer on line or the last in a chain.
User serial communications interface. Serial wiring
supplied by user.
4–20 mA output to user recording equipment. Wiring
supplied by user.
Connections for: Data Valid, Concentration Alarm, and
System Alarm. Wiring supplied by user.
PN 583054901, Rev YA