Advertisement

Sample Temperature and Pressure Requirements

Inlet sample temperature:
Optimum input gas temperature:
High Pressure
Minimum input pressure:
Maximum pressure:
Low Pressure (See specific gas table below)
Minimum input pressure:
Maximum pressure:
Helium
Hydrogen
Nitrogen
Oxygen
Argon
If necessary, a UHP pressure regulator can be installed at the main process
shut-off valve.
2-6 | 5830 Moisture Analyzer
For optimum performance, the sample line should be heat traced to
maintain a constant sample temperature.
0 °C to 100 °C (32 °F to 212 °F)
60 °C ± 5 °C (140 °F ± 9 °F)
138 kPa (20 PSIG)
345 kPa (50 PSIG)
34.5 kPa (5 PSIG)
138 kPa (20 PSIG)
5–20 PSIG (34.5–138 kPa)
5–20 PSIG (34.5–138 kPa)
5–20 PSIG (34.5–138 kPa)
5–20 PSIG (34.5–138 kPa)
7–20 PSIG (48.3–138 kPa)
PN 583054901, Rev YA