Mass Flow Meter; Removing The Mass Flow Meter - Ametek 5830 User Manual

Moisture analyzer
Hide thumbs Also See for 5830:
Table of Contents

Advertisement

Mass Flow Meter

Removing the Mass Flow Meter

1. Remove the Analyzer Cover.
2. Remove the Sample Oven Cover. Facing the front of the 5830 Analyzer, the
3. Disconnect the 1/16-inch fittings on all four (4) sample tubes, which are
4. Remove the four (4) nuts that secure the sample manifold plate (the square
5. Disconnect the cable from the Flow Meter (Figure 4-4).
6. Disconnect the 1/16-inch tubing connections at both ends of the Flow
7. From the bottom side of the sample manifold plate, remove the screws
PN 583054901, Rev YA
Mass Flow Meter is the small black box on the left-hand side of the Sample
Oven (Figure 4-2).
located on the bracket in the center of the Sample Oven (Figure 4-1).
aluminum mounting plate, located in the front section of the Oven), and
gently lift the manifold out of the Oven.
All of the wiring connections should still be intact at this time; you
will need to rest the sample manifold plate on top of the analyzer to
remove the Flow Meter.
Meter (Figure 4-4).
securing the Flow Meter (Figure 4-4). Remove the old Flow Meter, and set it
aside.
Maintenance and Troubleshooting | 4-7