Electrical Connections

Voltage Selection

The 5830 Analyzer comes from the factory set to operate on 120 VAC. If you
must switch the setting to 240 VAC (or back to the 120 VAC setting), follow the
procedure below.
To change the voltage setting (Figure 2-6):
1. Ensure that the area is non-hazardous and the analyzer has been de-ener-
2. Remove the Terminal Cover the analyzer Rear Panel.
3. Select 115 or 230 VAC on the Voltage Selection Switch (Item 10) located on
4. Replace Terminal Cover.
5. Clearly mark the operating voltage on the analyzer serial number label, in
Figure 2-6.
115 / 230 voltage selection
switch.
PN 583054901, Rev YA
This analyzer has an internal AC Voltage Selection Switch that must be
set correctly to prevent damage. Only qualified service personnel may
access the interior of the analyzer. If qualified service personnel are not
available, contact AMETEK Service.
You must follow this procedure to prevent damage to/malfunction of
the analyzer.
gized and power disconnected
the top left-hand side of the user User Interface board.
the lower right-hand corner of the analyzer Rear Panel.
LINE VOLTAGE
10
Voltage
Selection
FUSE
TB1
Switch
L
N
G
1
2
3
RS-485
TERMINATOR
TB2
ANALOG
OUTPUT
DATA
CONC
SYSTEM
RD
VALID
ALARM
ALARM
-
+
+
OUT
IN
1
2
3
4
5
6
7
8
9
10
11
12
Installation and Start-Up
ON
ON
TB3
TD
-
+
-
13
14
|
2-13

Advertisement

Table of Contents
loading

This manual is also suitable for:

583054901

Table of Contents