Production Codes; Moisture Unit - Ametek 5830 User Manual

Moisture analyzer
Hide thumbs Also See for 5830:
Table of Contents

Advertisement

Production Codes

Use this function to enter the production codes for the Moisture Generator
and Reference Dryer.
Figure 3-6.8.
Production Codes menu.
Generator
Reference Dryer

Moisture Unit

Figure 3-6.9.
Moisture Unit menu.
Select Unit
PN 583054901, Rev YA
Config
Production
Codes
The production code for the Moisture Generator is required for proper
operation of the analyzer. This number contains the nominal output of the
device, which is used to determine the analyzer span settings.
The dryer production code is required for proper operation of the analyzer.
This number contains the dryer capacity value, which is used to estimate
the life expectancy of the device.
Use caution, as entering a "new" value informs the analyzer that a
new dryer has been installed in the reference leg of the analyzer. The
remaining dryer capacity is reset when a new number is entered.
Config
Moisture Unit
Select Unit
The units for the moisture readings can be selected from a list. Pressing
Enter at the "Select Unit" prompt displays the list. (If a dew/frost point is
selected, you must enter a value for the process pressure in units of kPa.
If you change the moisture units, remember to make the necessary
changes to the Alarm and Analog Range settings.
Generator
Reference
Dryer
PPMV
PPBV
Pressure (KPa)*
Centigrade
Pressure (KPa)*
Fahrenheit
Controller/Interface
|
3-13

Advertisement

Table of Contents
loading

This manual is also suitable for:

583054901

Table of Contents