Installing The New Dryer - Ametek 3050-AMS User Manual

Moisture analyzer
Hide thumbs Also See for 3050-AMS:
Table of Contents

Advertisement

To Sample Dryer
Figure 7-6b.
Model 3050-AMS dryers,
connections.
7-8 | 3050 AMS Moisture Analyzer

Installing the new Dryer

Minimize the time that the dryer is exposed to the atmosphere after
!
removing the protective caps. AMETEK recommends that the dryer
CAUTION
be exposed for less than a minute.
1.
Insert the replacement dryer into the dryer port on the analyzer. Use
the new VCR gaskets provided with the dryer in each of the fittings.
2.
Tighten the VCR fittings until they are finger tight.
3
Install the dryer mounting screws to secure the dryer to the bracket.
4.
Tighten both of the VCR fittings so that the nuts are 1/8
finger tight.
5.
Place the dryer compartment cover onto the rear of the analyzer and
secure with the four captured screws.
6.
Enter the production code for the new dryer.
To Dryer
From Dryer
of a turn past
th