MCU Board

The MCU board (AMETEK Part No. 583068901S) is located below the Analog
Interface board (Figures 4-9 and 4-11).

Removing the MCU Board

Replacing the MCU Board

4-20 | 5830 Moisture Analyzer
Remove the Analog Interface board as described in the previous section.
Using a 1/16-inch hex wrench, remove the four (4) standoffs that secure
the MCU board in place. Unplug all connectors (Figures 4-11).
Install the new MCU board on the mounting plate using the four (4) stand-
offs removed and a 1/16-inch hex wrench.
Remount the Analog Interface board as described in the previous section.
PN 583054901, Rev YA