Variables Table - Ametek 5830 User Manual

Moisture analyzer
Hide thumbs Also See for 5830:
Table of Contents

Advertisement

Variables Table

Name
Press Units
MaxZeroError
MaxZeroDrift
AnalogOutSpan
AnalogOutOffset
Baud
SerialMode
NodeAddress
AnalyzerName
Gas
Alarm1Hi
Alarm1Lo
MaxCalError
KFactor
CalHour
CalPeriod
CalMonthDays
CalWeekDays
CalType
SerialNumber
CellSerialNumber
MoistureGeneratorSN
Version
HoldOut
VerDurationLimit
ZeroOnly
EnableAlarm1
DryerDateCode
DryerPpmHours
LCDcontrast
LCDbacklight
AdjustFreqOffset
ZeroHour
ZeroPeriod
ZeroMonthDays
ZeroWeekDays
ZeroType
Pressure1
CellTemp
Figure 2-22.1.
Variables Table.
2-28 | 5830 Moisture Analyzer
ID
HEX
Description
13
D
Process pressure units of measure
18
12
Maximum zero error allowed
19
13
Maximum zero drift allowed
29
1D
Analog output span
30
1E
Analog output offset
40
28
Serial communication speed
41
29
RS-485 two-wire or four wire mode
42
2A
RS-485 analyzer address
43
2B
Ametek analyzer name
44
2C
Sample gas name
46
2E
High concentration limit
47
2F
Low concentration limit
55
37
Maximum verification error allowed
56
38
Gas based coefficient
57
39
Verification schedule parameter
58
3A
Verification schedule parameter
59
3B
Verification schedule parameter
60
3C
Verification schedule parameter
61
3D
Verification schedule parameter
62
3E
Analyzer manufacturing number
63
3F
Sensor manufacturing code
64
40
Moisture generator production code
65
41
Embedded software version
69
45
Hold out during zero and verification
71
47
Minimum verification cycles number
72
48
Start zero only flag
73
49
Enable or disable concentration alarm
75
4B
Dryer production code
76
4C
Dryer consumption in 1000 ppm hours
79
4F
Display contrast setting
80
50
Display back light on or off
81
51
Adjust frequency offset after zero
82
52
Zero schedule parameter
83
53
Number of cycles to stay in zero
84
54
Zero schedule parameter
85
55
Zero schedule parameter
86
56
Zero type(daily,monthly,weekly)
87
57
Sensor pressure value
88
58
Sensor temperature value
Type
Typical Value
Integer
0
Float
0.5
Float
1.0
Float
1.0
Float
0.0
Integer
9600
Integer
1(four wire)
String
1
String
5830
String
Nitrogen
Float
100.0
Float
0.0
Float
unused
Float
1.0
Integer
13
Integer
15
Integer
1
Integer
1
Integer
0
String
123
String
583015
String
020101305012
String
1.00
Integer
1(hold)
Integer
15
Integer
1(zero only)
Integer
1(enable)
String
010080009
Float
20000.00
Integer
5
Integer
1(on)
Integer
1(adjust)
Integer
13
Integer
15
Integer
1
Integer
1
Integer
0(daily)
Float
14.7 psi
Float
60.0 degreeC
PN 583054901, Rev YA