Sensor

Removing the Sensor

1. Remove the Analyzer Cover.
2. Remove the cover from the Sample Oven Cover. Facing the front of the
3. Disconnect the two (2) 1/16-inch fittings that connect the Sensor to the
4. Lift the Sensor assembly from the Oven compartment being careful to free
5. Disconnect the ribbon cable from the Analog Interface board and discard

Installing the Sensor

1. Mount the new Sensor on the stream switching valve manifold by carefully
2. Run the ribbon cable from the Sensor through the slotted opening in the
3. Connect the ribbon cable plug to position P2 on the Analog Interface
4. Replace the Sample Oven Cover and secure it with the two (2) nuts.
5. Replace the Analyzer Cover and secure it with the mounting screws.
4-6 | 5830 Moisture Analyzer
5830 Analyzer, the Sensor is the small PC board with two (2) metal cans in
the center of the sample manifold plate (Figure 4-2).
stream switching valve manifold in the center of the sample manifold plate
(Figure 4-3).
the ribbon cable connected to it from the other wires that pass through to
the electronics compartment.
the old Sensor.
inserting and tightening the two (2) 1/16-inch fittings.
Oven wall into the electronics compartment.
board (Figure 4-10).
Use caution when disconnecting the Moisture Generator and pressure
transducer from the Sample System. The tubes are delicate and care
should be taken not to transfer torque to them.
PN 583054901, Rev YA