Overview - Ametek 5830 User Manual

Moisture analyzer
Hide thumbs Also See for 5830:
Table of Contents

Advertisement

The 5830 Moisture Analyzer reliably measures low concentrations of moisture
in high purity gases at concentrations down to 20 parts-per-billion by volume
(PPBV), using a quartz crystal microbalance (QCM) sensor. The calibrated range
of the analyzer is 0 to 100 PPMV with a trend indication for measured concen-
trations to 1000 PPMV. The QCM-sensor technology used in this analyzer is
compatible with inert gases, air, oxygen, nitrogen, hydrogen, Freons (12, 22,
32, 114, 116, 125, 134A, and 152), methane, ethane, propane, and some spe-
cialty gases (e.g., sulfur hexafluoride). Carbon dioxide requires a custom Sensor
assembly. Please contact the factory to confirm compatibility with other gases.
Figure 1-1.
5830 Moisture Analyzer.
PN 583054901, Rev YA

Overview

|
Overview
1-1