Electrical Requirements

The 5830 Analyzer must be installed in a general purpose area. It does not
have an On/Off power switch so power must be applied through a circuit
breaker switch at the AC source.
General Purpose Installation:

Instrument Air Supply

PN 583054901, Rev YA
Power connection can be made using the supplied power cord(s) termi-
nated with a locally approved wall socket plug, or
The analyzer can be hard-wired.
Use dry instrument air (70–100 PSI) that is free of oils and other contami-
nants for valve operation. N
If there is any doubt about the cleanliness of the air supply, connect it
through oil / air separators.
Air tubing must be fitted to a 1 /4-inch Swagelok® fitting.
can be used as an alternate.