Verify Schedule

Use this function to schedule automatic analyzer Verification.
Figure 3-6.7.
Production Codes menu.
Schedule Type
Day of the Week
Day of the Month
Hour
Duration
3-12 | 5830 Moisture Analyzer
Config
Verify Schedule
Choose the preferred Schedule Type:
Never, Daily, Weekly, or Monthly.
Select the day of the week from the list.
Enter the 2-digit value for the day of the month.
Enter the 2-digit value for the hour of the day using a 24-hour clock. Zero is
the starting hour.
Enter the Zero period value in minutes.
AMETEK recommends performing a Verification cycle every week for
the first month of operation. After this start-up period, the analyzer
will have dried down sufficiently to increase the Verification interval
to 1 month. AMETEK recommends that the duration of the Verification
cycle be set to 60 (i.e., 60 minutes).
Never
Daily
Weekly
Monthly
Schedule
Type
Sunday
Monday
Tuesday
Day of
Wednesday
Week
Thursday
Friday
Saturday
Day of
Month
Enter Value
Enter Value
Hour
Enter Value
Duration