Dryer Installation - Ametek 3050-AP User Manual

Moisture analyzer
Table of Contents

Advertisement

Dryer Installation Instructions
Tools needed to install dryer:
One common flat-head screwdriver
Two 7/16-inch wrenches
One 5/8-inch wrench
One 3/4-inch wrench
1. To alleviate stress on the tube connections attach dryer to dryer bracket using two 10-32
screws, flat and lock washers. Refer to figure 2.4
2. Install adapter tubes into the dryer input and output on the analyzer.
a. Assemble loose female VCR nuts and VCR glands.
b. Remove plastic dust caps from the dryer input and output on the analyzer.
c. Remove compression nuts and verify ferrule is properly installed.
d. Reinstall nuts to finger tight.
e. Insert adapter tubes through female compression nuts and ferrules until they
bottom out.
f.
Using two 7/16-inch wrenches (one to hold the bulkhead fitting), swage the
compression fittings into the bulkhead fittings on the analyzer. For proper
installation, the fitting should be
3. Attach dryer to adapter tubes.
Do not leave the dryers open to ambient air. The high moisture levels present
!
in ambient air can damage or shorten the life span of these components. Make
dryer connections promply.
CAUTION
a. Remove female VCR nuts from dryer VCR glands.
b. Verify VCR gaskets are in place.
b. Align adapter tubs with dryer VCR glands.
c. Slide female nuts over VCR adapters and screw nuts onto dryer glands.
Secure VCR nut behind VCR adapter on dryer with 5/8-inch wrench. Use a
3/4-inch wrench to tighten nuts
d. Check for leaks.
2-6 3050-AP Moisture Analyzer
/
turn past finger-tight.
1
4
/
turn past finger-tight.
1
8

Advertisement

Table of Contents
loading

Table of Contents