Pilot Valve Manifold; Removing The Pilot Valve Manifold - Ametek 5830 User Manual

Moisture analyzer
Hide thumbs Also See for 5830:
Table of Contents

Advertisement

Pilot Valve Manifold

The Pilot Valve Manifold (AMETEK Part No. 583011901S) is found on the side of
the Power Supply PCB Board (Figures 4-9).

Removing the Pilot Valve Manifold

1. Using a 7/64-inch hex wrench, loosen the two (2) captured screws that
2. Disconnect the pilot valve receptacle plugs (Figure 4-9).
3. Use a 1/16-inch slotted screwdriver to remove the two (2) screws at the
4. Detach the tubing from the manifold by holding the tubing retaining col-
Figure 4-12.
5830 pilot valve manifold.
4-22 | 5830 Moisture Analyzer
secure the Pilot Valve Manifold to the sample manifold plate (Figure 4-9).
side of the manifold (Figure 4-12).
The mounting screws for the two manifolds are different size.
lar and pulling the tubing free. Make sure that you identify the tubing for
proper position when replacing.
Receptacle Plug
Screws
(Use Slotted
Screwdriver)
PN 583054901, Rev YA