Analyzer Installation; User-Supplied Items; Interconnect Wiring; External Sample Tubing - Ametek 5830 User Manual

Moisture analyzer
Hide thumbs Also See for 5830:
Table of Contents

Advertisement

Analyzer Installation

User-Supplied Items

Interconnect Wiring

The user must supply all wiring for serial communication, Analog Output, and
alarm relays. Recommended wire is 22 gauge, twisted/shielded pair.

External Sample Tubing

PN 583054901, Rev YA
Consult plant safety personnel for appropriate exhaust venting guide-
lines for specific Sample gas type.
Sample input tubing length should be as short as possible.
-
For optimum performance, sample tubing should be thermally con-
trolled to maintain a constant sample input temperature. Optimal
results are obtained with the sample line maintained at 60 °C.
-
Recommended Sample Inlet fitting tubing is 1/4-inch OD, electro pol-
ished 316L VAR stainless steel (AMETEK Part No. 257707002 or equiva-
lent).
Exhaust pressure must be 0–15 PSIG.
Input and exhaust tubing must be fitted with female 1/4-inch VCR connec-
tors.