Write Data (H)
The data field of this frame contains the handle for a system variable. See Fig-
ure 2-22.1 and 2-22.2 for a list of defined handles for system variables.
Example:
Master:
Slave:
Write Cell EEPROM (L)
Copy cell serial number and cell calibration coefficients into cell EEPROM.
Example:
Master:
Slave:
Quit (Q)
Zero cycle is terminated immediately and normal operation resumed.
Example:
Master:
Slave:
Zero (Z)
Initiates a Zero cycle.
Example:
Master:
Slave:
2-26 | 5830 Moisture Analyzer
>34H05123.40C[CR]
A[CR]
>34LB3[CR]
A[CR]
>34QB8[CR]
A[CR]
>34Z??[CR]
(Note the use of wildcard in place of checksum)
A[CR]
PN 583054901, Rev YA