Installing the New Moisture Generator

1. Install the new Moisture Generator on the mounting bracket in the correct
2. Carefully reconnect the 1/16-inch fittings (two on the generator and one
3. Place the sample manifold plate back in the Oven and secure with the four
4. Reconnect all four (4) 1/16 fittings on the bracket in the center of the Oven.
5. Replace the Sample Oven Cover, and secure it with the two (2) nuts (Figure
6. Replace the Analyzer Cover and secure it with the mounting screws.
7. Power up the analyzer. Wait until the boot-up sequence is complete. Enter
PN 583054901, Rev YA
Trans ducer
Moisture
Generator
orientation (Figure 4-6) and remount the bracket on the sample manifold
plate.
on the pressure transducer).
(4) mounting nuts.
4-1).
the new production code in the software (see details in the Variables
table in Chapter 2, and refer to "Production Codes" and "Working From the
Schedule Tab" in Chapter 3).
Pres sure
Maintenance and Troubleshooting | 4-11
Mani f ol d