5-136 Menus and Commands

SelfTest Dialog

Path: Main Menu Service Menu Diags
Submenu SelfTest Dialog
Use the SelfTest Dialog to run a predetermined
sequence of diagnostics and exercisers. See Figure 5-187.
The selftest runs
a sequence of diags.
Cycles to run:
Figure 5-187
Parameter
Cycles to Run
Accept
If Y is selected for the Accept parameter value, the
Response Dialog appears. See Figure 5-188. Otherwise,
the changed parameters continue to display but no action
is taken.
DIAGS SELF TEST
Cycle :
Status:
Figure 5-188
Parameter
Cycle
Status
Cancel
009<
Accept: N

SelfTest Dialog

Value
Indicates the desired cycle
count (001–999).
Y to accept changes
N to reject changes
006 of 009
Running
Cancel: N<
Response Dialog
Value
The number of completed test
cycles of the requested cycles.
Shows each test as it runs
(Loopback, Sensors, Lock,
Unlock, Open, Close, Extend,
Retract, Move, Scanner), then
displays the status of the
indicated cycle (Completed,
Error, Canceled).
Y to cancel the test
N to continue running the test