5.7 Dynamic Trial Fitting
CAUTION
Adjusting the Settings
Fall due to incorrect or unfamiliar settings
Only adapt the settings to the patient gradually.
Explain the effects of the adjustments on the use of the prosthesis to the patient.
INFORMATION
Established initial tension values are used for the settings of the prosthetic knee joint at the fact­
ory.
If a reduction of the initial tension is required, ensure that the initial tension is always free of
play.
When initial tension for a setting is reduced to a minimum, ensure constant freedom of play
for the setting (e.g. by establishing a maintenance interval to verify the adjustment values).
This is because the adjustment values can change during use of the prosthesis.
During dynamic trial fitting, the alignment and settings of the prosthesis are checked and adapted
for optimum walking according to the needs and abilities of the patient.
The patient has to learn the safe use of the prosthesis through intensive training.
The following subsections describe the adjustment possibilities for adapting the product to the
patient.
The list below provides an overview of the sequence for working through the subsections:
Checking the Factory Settings
Adjusting the EBS
Adjusting the extension assist
Adjusting the swing phase friction
Fine-tuning the settings
5.7.1 Checking the Factory Settings
Function
EBS
Extension assist
Swing phase friction
1) CAUTION! Do not adjust the factory settings prior to the dynamic trial fitting.
Note the position of the adjusting screw
tial position can be restored if required. Use the markings on the adjusting screw and the
plastic cover
Means of adjustment
Adjusting wheel
(see fig. 11 -
)
Adjusting wheel
(see fig. 13 -
)
Adjusting screw
(see fig. 15 -
)
for this purpose (see fig. 15).
Factory setting
3R62, 3R62=KD,
3R62=ST:
Spring
with
adjusting
wheel, 1.5 mm initial ten­
sion
3R62=1, 3R62=1-KD,
3R62=1-ST:
Spring
with
adjusting
wheel, 1.0 mm initial ten­
sion
Spring
with
adjusting
wheel, 1.5 mm initial ten­
sion
Adjusting
tightened to 2.0 Nm
upon delivery as well as all changes so that the ini­
Meaning
Low initial tension as the
starting point for adjust­
ment to the patient
Low initial tension as the
starting point for adjust­
ment to the patient
screw
Standard braking effect
as the starting point for
adaptation to the patient
Ottobock | 35