3) Screw the threaded fitting
4) Screw the lock slide
5) Verify the proper fit of the threaded fitting in the lock slide.
6) CAUTION! The lock and the lock slide affect patient safety and must function reliably
in all situations.
Verify the functionality of the lock and the lock slide at maximum extension and flexion of the
prosthetic knee joint.
5.4 Information on Using a Cosmetic Cover
CAUTION
Use of talcum
Falling, damage to the prosthetic knee joint due to lack of lubrication
Do not use talcum on the prosthetic knee joint or other prosthetic components.
Inform the patient.
► Only use the foam cover approved by the manufacturer (see section "Combination Possibilit­
ies" – see Page 24).
► To reduce friction and to eliminate noise, apply 519L5 Silicone Spray directly onto the contact
surfaces of the cosmetic foam cover.
► After fabricating the cosmetic cover, repeat the dynamic trial fitting since the cover affects the
adjustments of the prosthesis (e.g. damping).
5.5 Bench Alignment
+ = anterior placement / – = posterior placement (in relation to the alignment reference line)
No.
see fig. 8
Required equipment:
Alignment apparatus (e.g., 743L200 L.A.S.A.R. Assembly or 743A200 PROS.A. Assembly),
743A80 50:50 Gauge, 743S12 Heel Height Measuring Apparatus
Position the prosthetic foot in the alignment apparatus in accordance with the following values:
   
Heel height:
effective heel height (x) + 5 mm
   
a–p positioning, middle of the prosthetic foot to the sagittal alignment reference line
+30 mm
Exterior foot rotation:
approx. 7°
m–l positioning, middle of the prosthetic foot to the frontal alignment reference line
0 mm medial
   
Check the prosthetic knee joint according to the section "Checking the Factory Settings" (see
Page 35).
Position the alignment reference point
joint in the alignment apparatus according to the following values:
Height of the alignment reference point in the alignment apparatus:
MTP-floor measurement + 20 mm
a–p positioning of the alignment reference point to the alignment reference line:
0 mm
m–l positioning, centre of the prosthetic knee joint to the frontal alignment reference line
0 mm
External rotation of the prosthetic knee joint:
approx. 5°
30 | Ottobock
onto the end of the perlon cable.
to the lamination disc
Bench alignment process
with the help of the pad button
(proximal, anterior rotation axis) of the prosthetic knee
.
:
:
: