Replacement Procedure

LCD Assembly Replacement

LCD Assembly Replacement

Tools Required

Removal Procedure

Refer to Figure 4-23 for this procedure.
Step 1. Remove the Outer Cover as described in "Outer Cover Replacement" on page 54.
Step 2. Remove the Front Panel as described in "Front Panel Replacement" on page 56.
Step 3. Remove the Touch Screen Control Board as described in "Touch Screen Control Board
Replacement" on page 89.
Step 4. Remove the Inverter Board as described in "Inverter Board Replacement" on page 90.
Step 5. Disconnect the two connectors (item 1) of the cables.
Step 6. Release the clamped cable from the clamp (item 2) on the LCD Assembly.
Step 7. Remove the two cables (item 3) on the LCD Assembly.
Step 8. Remove the four TORX T15 screws (item 4) fastening the LCD Assembly.
Step 9. Lift the LCD Assembly carefully.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
Set a torque to 14 in-lb for torque driver when fasten the TORX T15 screws (item 4).
96
torque screwdriver, T8, T10 and T20
torque screwdriver, T15 (set to 14 in-lb)
flat edge screwdriver