To fit a borosilicate glass capillary (nanovial):
Warning:
contamination with biohazardous and toxic samples, do not touch
the sharp end of the capillary.
To avoid damaging capillaries, take great care when handling
Caution:
them; they are extremely fragile. Always hold their blunt end, never the
sharp end.
To avoid electric shock, ensure that the NanoFlow stage is
Warning:
fully retracted from the source before beginning this procedure.
1.
Loosen the stage retaining screw.
2.
Pull the stop screw to release the stage.
3.
Slide the stage out of the NanoFlow source enclosure, and remove the
magnetic cover.
4.
Unscrew the retaining screw, and lift the sprayer from the stage.
5.
Unscrew the union from the end of the sprayer assembly.
6.
Remove the existing capillary from the sprayer.