Waters Xevo TQ-S Operator's, Overview And Maintenance Manual page 153

Mass spectrometry system
Table of Contents

Advertisement

3.
Use the 3-mm Allen wrench to unscrew and remove the 4 screws
securing the PEEK ion block support to the adaptor housing.
Adaptor housing
PEEK ion block
support
Securing screws
4.
Remove the PEEK ion block support from the adaptor housing.
5.
Use the O-ring removal kit to carefully remove all the O-rings from the
PEEK ion block support (see
6.
If any of the O-rings show signs of deterioration or damage, dispose of
them in accordance with local environmental regulations.
Cleaning the StepWave ion guide assembly
page
110).
To avoid contaminating uncontaminated
Warning:
surfaces with biohazards or compounds that are toxic
or corrosive, dispose of the O-rings according to local
environmental regulations.
February 28, 2014, 715002212 Rev. C
StepWave
assembly
153