3 Changing the Mode of Operation
¼-inch wrench
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while performing this procedure. The source
components can be contaminated.
Warning:
performed on its source before beginning this procedure.
To remove the ionKey source:
1.
Prepare the instrument for working on its source (see
Warning:
with the instrument's source enclosure open; the source can be hot.
2.
Remove the iKey from the docking port (see the ionKey/MS System
Guide).
3.
Shut down the MassLynx software.
4.
Power-off the µSample manager.
5.
Disconnect the PSPI cable.
6.
Using the ¼-inch wrench, loosen and disconnect the fluid waste line and
fluid inlet lines from the µSample manager.
7.
Swing open the ionKey source enclosure unit from the source mounting
on the mass spectrometer.
8.
Disconnect the high voltage cable (white) from the high voltage supply
outlet on the mass spectrometer.
9.
Disconnect the reference probe power cable (green) from the reference
probe power inlet on the mass spectrometer.
10. Disconnect the fluid infusion line from the onboard IntelliStart Fluidics
on the mass spectrometer.
11. Carefully lift off the ionKey source module and store safely.