Positioning The Borosilicate Glass Capillary Tip - Waters Xevo TQ-S Operator's, Overview And Maintenance Manual

Mass spectrometry system
Table of Contents

Advertisement

Sprayer Assembly:
5 mm
12. Screw the sprayer back into the assembly.
13. Replace the sprayer cover.
14. On the MassLynx MS Tune window, ensure that the Capillary
parameter on the ES+/- Source tab is set to 0 kV.
Caution:
tip position before you push the sprayer inside the NanoFlow
source enclosure. Ensure that the capillary tip does not collide
with the cone or the side of the source.
15. Carefully push the stage back into the NanoFlow source enclosure, using
the stop and handle.

Positioning the borosilicate glass capillary tip

Having obtained a signal, you must adjust the tip position to maximize it.
Using the three-axis manipulator, you can adjust the tip position up and
down, left and right, forward and backward. As a starting point, set the tip so
that it is on the center line of the sampling cone and at a distance between two
PTFE tubing
Ferrule
Union
Knurled nut
Blue conductive elastomer
Glass capillary
To avoid damage to the capillary tip, adjust the sprayer
February 28, 2014, 715002212 Rev. C
NanoFlow ESI source
81

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents