Waters Xevo TQ-S Operator's, Overview And Maintenance Manual page 221

Mass spectrometry system
Table of Contents

Advertisement

Water
Isopropanol or methanol
Source of clean, pressurized air
To avoid damage, do not use abrasives or solvents to clean the
Caution:
instrument's case.
During normal operation, the ionKey source does not require cleaning. After
repeated use, however, material can accumulate around the fluid-line
connectors or electronic connectors in the iKey docking port.
Cleaning the fluid-line connectors and electronic connectors:
To remove buildup from fluid-line connectors:
Gently swab the fluid-line connectors and outer edges of the docking port with
a lint-free tissue saturated with isopropanol or methanol.
To remove buildup from electronic connectors:
1.
Open the source door.
2.
From the back, gently direct a stream of air from a clean source across
the electronic connectors.
To prevent contamining the inside of the source, do not direct air
Note:
into the source.
3.
Close the source door.
Cleaning the ionKey source and connectors
Electronic connectors
Fluid-line connectors
February 28, 2014, 715002212 Rev. C
221