3 Changing the Mode of Operation
For further details, see
The following sections explain how to install and remove an ESI probe.
Installing the ESI probe
Required materials
•
Chemical-resistant, powder-free gloves
•
Sharp knife or PEEK tubing cutter
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while performing this procedure. The LC
system connections, ESI probe, and source can be contaminated.
Warning:
prepare the instrument according to the procedure on
To install the ESI probe:
1.
Prepare the instrument for working on the source (see
Warning:
care; the probe tip is sharp.
2.
Remove the protective sleeve, if fitted, from the ESI probe tip.
58
February 28, 2014, 715002212 Rev. C
"ESI" on page
32.
To avoid personal contamination with biohazards or
To avoid electric shock, before beginning this procedure,
To avoid puncture wounds, handle the ESI probe with
page
98.
page
98).